3A.36

Correction of a varus deformity of the distal femur which has occurred during lengthening with the LRS ADVanced.

(a) Angular deformity following lengthening procedure; the distal clamp of 53000 series fixator is a Micrometric Swivelling Clamp.

(b) Deformity corrected by acute shortening and callus manipulation.

Use of HA-coated screws is recommended.

Note: When a Paediatric Swivelling Clamp is sed, it should be applied only for correction of the deformity. Do not lengthen with the Paediatric Swivelling Clamp.

Equipment required
Q.tyCatalogue N.Description
1
53515R
ADV RADIOLUCENT LRS KIT 200MM
1
53115
ADV MICROMETRIC SWIVELLING CLAMP
4
99-60165
OSTEOTITE BONE SCREW 150/40 MM SHAFT D 6 MM THREAD D 6-5 MM
1
99-60102
OSTEOTITE BONE SCREW 150/50 MM SHAFT D 6 MM THREAD D 6-5 MM
1
99-60111
OSTEOTITE BONE SCREW 150/60 MM SHAFT D 6 MM THREAD D 6-5 MM
Operative techniques